SECTION 5

– PHOTOMETRICS

The photometrics measurement mode enables simple measurements of absorbance, %

transmittance and concentration to be performed. In this measurement mode it is possible to calibrate

against a standard of a known concentration or use a known factor. The sample is measured at one

wavelength at one point in time. There are no post measurement calculations available in this

measurement mode.

5.1

PHOTOMETRICS MENU SCREEN

The photometrics measurement mode screen enables

measurement parameters to be changed. The four

options at the bottom of the screen allow the user to

select the unit of measurement, the measurement

mode, the concentration factor and the calibration

standard

’s concentration. These options are selected

with the corresponding function buttons.

5.2

METHOD SET UP

The photometrics measurement mode is very simple and the only parameters which can be adjusted

are the wavelength, the units of the measurement the measurement mode and the concentration

setup parameters.

5.2.1

Selecting a Wavelength

The wavelength can be adjusted by pressing the GOTO

key on the instrument keypad. The required

wavelength can then be entered using the numerical keypad. Press the Enter key to confirm.

5.2.2

Selecting the Measurement Mode

The measurement mode can be selected by pressing the function key below the Mode option. The

required measurement mode can then be selected from the available options of Abs, %T and

Conc/Factor using the up and down arrow keys. Press the Enter key to confirm.