Deuterium lamp replacement, D2 lamp tungsten lamp switch mirror – Jenway 6850 User Manual

Page 53

Advertising
background image

53

3. Whilst wearing cotton gloves remove the

defective tungsten lamp and replace with a

new one.

4. Switch on the power and move the Switch

Mirror down into the position shown in the

picture opposite. Observing the entrance

window

on the instrument’s monochromator,

adjust the tungsten lamp until the light is

centred on the entrance hole.

5. Disconnect the power and re-attach the lamp

compartment cover and the top case of the

spectrophotometer.

12.2.2 Deuterium Lamp Replacement

1. Unscrew

the

4

screws

on

the

spectrophotometer’s top case (see 12.2.1)

and

remove

the

cover

of

the

spectrophotometer.

2. Unscrew the 2 screws on the top of the lamp

chamber and remove the cover. (see 12.2.1)

3. Unscrew the 2 screws on the deuterium lamp

flange (No.1), unplug the connector on the

power board (No. 2) and remove the

deuterium lamp. Whilst wearing cotton gloves

replace with a new deuterium lamp. Re-fix the

2 screws and reconnect the power cable.

4. Re-attach the lamp compartment cover and

the top case of the spectrophotometer.

D2 Lamp

Tungsten

Lamp

Switch

Mirror

Entrance Window

Entrance

Hole