7.6

SPECTRUM SEARCH

The photometric values of the measured spectrum can

be viewed by pressing the function key below the

Search option. The user can press the up and down

arrow keys to move between the peaks in the scan that

are above the threshold peak height. The user can

alternatively press the left or right arrow keys to review

each point in the scan in turn.

The peak height threshold can be adjusted by pressing

the function key below the PK Height option. The

threshold peak height value can be entered using the

numeric keypad. Press the Enter key to confirm.

7.7

SPECTRUM SMOOTHING

The measured spectrum scan can be smoothed by

pressing the function key below the Smooth option.

7.8

STORING RESULTS

A result file can b

e saved to the instrument’s internal

memory.

Press the Save key on the instrument keypad. The

instrument will ask the user to enter a filename (max. 8

characters) using the alpha-numeric keypad. The user

can select the indicated letters, numbers and symbols

by repeatedly pressing the key until the required value

is displayed, after a short period the next character can

be entered in the same manner. Once the filename has

been entered, press the Enter key to confirm.