CREATING A NEW STANDARD CURVE

The user must firstly edit the calibration table before attempting to create a new standard curve.

Insert the cuvette containing the blank solution into the

reference position in the sample chamber and insert

the cuvette containing the first standard solution into

the sample position. Close the instrument lid and press

the start/stop button. Once the measurement is

complete the instrument will ask for the second

standard solution to be inserted. Repeat the

measurement procedure for each standard solution.

Once all standard solutions have been measured the

new standard curve and curve equation will be shown.

The standard curve can be viewed graphically by

pressing the function key below the show curve option

on the calibration table screen.