6.3.2.2

Manually Enter Curve Constants

The user can manually enter the calibration curve

constants by pressing the function key below the

Params option. The user will be prompted to enter the

constants that are required for the selected curve fit

algorithm.

1. Linear Fit = K0 + (K1 x Abs)

2. Square Fit = K0 + (K1 x Abs) + (K2 x Abs

2

)

3. Cubic = K0 + (K1 x Abs) + (K2 x Abs

2

) + (K3 x

Abs

3

)

4. Linear Fit Through Zero = K1 x Abs

The values can be entered using the numerical keypad.

Press the Enter key to confirm.

6.3.2.3

Edit Calibration Table

The user can edit the calibration curve data by pressing

the function key below the Standard option. The user

will be prompted to enter the concentration values for

each standard. The values can be entered using the

numerical keypad. Press the Enter key to confirm.

To add additional standard values to the table, press

the down arrow on the keypad. Once all values have

been entered, press the Esc key on the keypad to

return to the calibration table settings screen.

6.3.2.4

Display Calibration Curve

Completed calibration curves can be viewed graphically by pressing the function key below the Show

curve option.