Section 1, Introduction, Instrument description – Jenway 6850 User Manual

Page 8: Instrument specification

Advertising
background image

8

SECTION 1

– Introduction

1.1

INSTRUMENT DESCRIPTION

The 6850 is a variable bandwidth, double beam UV/visible spectrophotometer with an integrated user

interface for local control. This instrument has highly stable optics and two detectors that measure the

sample and reference solutions simultaneously to optimise the accuracy of the measurement. The

6850 has measurement modes for photometrics, concentration, multi-wavelength, spectrum

scanning, quantitation, kinetics and DNA and Protein analysis. The instrument is supplied with

Jenway Prism PC software to allow full PC control of the spectrophotometer.

1.2

INSTRUMENT SPECIFICATION

6850

Wavelength
Range

190 to 1100nm

Resolution

0.1nm

Accuracy

± 0.3nm (at 0.5 and 1nm bandwidth) ± 0.5nm (at 2, 4 and 5nm

bandwidth)

Repeatability

± 0.2nm

Spectral bandwidth

Variable 0.5 / 1 / 2 / 4 and 5nm

Photometrics
Transmittance

0 to 200%

Absorbance

-0.3 to 3.0A

Accuracy

±0.3%T (0

– 100%T), ±0.002A (0 – 0.5A)

Reproducibility

±0.001 Abs (0 to 0.5 Abs) ±0.002 Abs (0.5 to 1.0 Abs)

Resolution

0.1%T, 0.001A

Stray light

<0.05% at 360nm and 220nm

Noise

0.0005A

Stability

±0.001A at 500nm after 15min warm up

Quantitation
Range

0 to 99999

Data points

Up to 3 wavelengths

Calibration

Blank with up to 10 standards

Units

IU, mM/L, M/L, μg/mL, mg/mL, mg/L, mEq, ppb, ppm, % and other

Curve fit algorithms

Linear, Linear through zero, Quadratic and Cubic

Multi-wavelength
Range

0 to 99999

Data points

Up to 10 wavelengths

Kinetics
Measurement Time

Up to 12 hours

Scan Interval

0.1, 0.2, 0.5, 1, 2, 5, 10 or 30 seconds

Calibration

Blank with a single standard or factor

Display

Graphical and concentration

Analysis

Slope and line of best fit between any two points

Spectrum
Range

Any range between 190 to 1100nm

Scan speed

100 to 2000nm/min

Scan interval

Selectable 0.1, 0.2, 0.5, 1, 2 or 5nm

DNA
Measurement
modes

DNA/RNA Ratio, concentration, A320 correction