Section 9, Dna/protein, Dna menu options – Jenway 6850 User Manual

Page 41: Adjusting the measurement mode

Advertising
background image

41

SECTION 9

– DNA/PROTEIN

The DNA/Protein measurement mode allows the user to measure multi-wavelength absorbance

ratios, such as 260nm/280nm and 260nm/230nm, which are commonly used to estimate a protein or

nucleic acid

sample’s purity. The mode also includes calculations that can be used to estimate the

concentration of the protein or nucleic acid sample.

9.1

DNA MENU OPTIONS

The

DNA/Protein

measurement

mode

enables

measurement parameters to be changed. The four

options at the bottom of the screen allow the user to edit

the concentration calculation factors, select the

measurement

mode,

select

the

units

of

the

measurement and reset the modes settings to their

default values. These options are selected with the

corresponding function buttons.

9.2

ADJUSTING THE MEASUREMENT MODE

The DNA/Protein measurement mode includes two sets

of calculations to calculate the concentration of nucleic

acid and protein samples. The measurement mode

option can be selected by pressing the function key

below the Mode option. The options that are available

are:-

1. Absorbance Difference 1

2. Absorbance Difference 2

The default equations and factors used by the

instrument are shown below:-

Absorbance Difference 1

Purity Ratio = Abs@260nm / Abs@280nm

DNA

Conc (μg/ml) = (Abs@

260

x 62.9)

– (Abs@

280

x 36.0)

Protein Conc (μg/ml) = (Abs@280 x 1553) – (Abs@260 x

757.3)