0 0.00h, Pressure, Zero adjustment procedure – KANOMAX 6501 series Climomaster User Manual

Page 16: Connecting the pressure tube

Advertising
background image

15

Pressure

Do not expose the pressure sensor to more than 75kPa of pressure.

Excess pressure may cause serious damage to the sensor.

The operating temperature must be between 5

o

C and 40

o

C or 41

o

F

and 104

o

F when measuring pressure. The instrument may not

operate properly outside of this temperature range.

Make sure to perform a zero adjustment before measuring

pressure. When performing a zero adjustment, leave both pressure

ports (+) and (-) open.

The zero adjustment range is – 0.5 to 0.5 kPa. If the reading is out

of this range, an error message will be displayed.

Zero Adjustment Procedure

Press

to enter the Main Menu.

Use

to select “8. PRESSURE ZERO”

and press

.

Use

to select “1. ZERO ADJUST”

and press

.

Use

to select “YES”.

Press

to display 0.00 kPa.

Use

to select “2. SAVE INFO”.

Press

. This will complete the zero

adjustment and you are back in Main
Menu.

If you press

before you save, you will go back to the Main Menu

without completing the zero adjustment.

Connecting the Pressure Tube

Connect the pressure tube to the (+) or (-)

port as you see on the left.
Connect/insert the other end of the tube to the
measuring hole of the duct. If the pressure
that you are going to measure is positive,
connect the tube to (+) port. If it is negative,
connect the tube to (-) port.

aaaa

PRESSURE

a

ZERO

1.ZERO

a

ADJUST

aaa

YES




2

.SAVE

a

INFO

0

0.00h

aaaa

PRESSURE

a

ZERO

1.ZERO

a

ADJUST

aaaY

ES





2.SAVE

a

INFO

0

0.00h

aaaa

PRESSURE

a

ZERO

1

.ZERO

a

ADJUST

aaa

YES





2.SAVE

a

INFO

-0.00h

MENU

aa

1.NORMAL

aaaaaa

2.DUCT

a

TYPE

aaaaaa

3.CALCULATION

aaaaaa

4.FLOW

a

RATE

aaaaaa

5.DATA

a

OUTPUT

aaaaaa

6.DATA

a

CLEAR

aaaaaa

7.UTILITY

aaaaaa8

.PRESSURE

a

ZERO

*

In order to take

an accurate
measurement,
make sure that the
tube is properly
secured without
any leakage, or
bends.

Pressure sensor is
only available on
MODEL6501-B0/-C0.

Advertising