Specification, 12 specification, 52 specification – KANOMAX 6501 series Climomaster User Manual

Page 53

Advertising
background image

52 Specification

§ 12

Specification

Product

CLIMOMASTER Air Velocity Meter

Model No. of the Main Body

6501

Model No. of the Probe

6531-21 6541

-21

6561

-21

6542

-21

6551

-21

6552

-21

6533-21

6543

-21

Measuring Object

Clean air flow

Air Velocity

[m/s]

Measuring Range

0.01 to 30.0, 0.01 to 50.0 (6561-21 only)

0.01 to 5.00

Resolution

0 to 9.99: 0.01

10 to 30: 0.1

30 to 50: 0.1 (6561-21 only)

0.01

Accuracy

±2% of reading or ±0.015 (±3FPM),

whichever is greater

0.01 to 0.99:±0.02

0.99 to 5.00:

±2% of reading or

±0.015 (±3FPM),

whichever is greater

Response Time

[sec.]

Approx. 1

Approx. 4

Approx. 7

Approx. 7

(at 1m/s, 90% response time)

Air

Temperature

[

o

C]

Measuring Range

-20.0 to 70.0

-20.0 to 70.0

Resolution

0.1

0.1

Accuracy

±0.5

±0.5

Response Time

[sec.]

Approx. 30

(at 1m/s, 90% response time)

Approx. 30

(at 1m/s, 90%

response time)

Humidity

*1

[%RH]

Measuring Range 2.0 to 98.0

2.0 to 98.0

Resolution

0.1

0.1

Accuracy

2 to 80:

±2.0

80 to 98:

±3.0

2 to 80:

±2.0

80 to 98:

±3.0

Response Time

[sec.]

Approx.

15

Approx.

15

Pressure

2

[kPa]

Measuring Range

-5.00~+5.00

Resolution

0.01

Accuracy

±3% of reading or ±0.01

Response Time

[sec.]

Approx. 1

Temperature Compensation

[m/s]

±5% of reading or ±0.015 (±3FPM), whichever is greater

(in the temperature range of 5 to 60.0

o

C)

Functions

Readings Hold, Max Hold, Time Constant (1, 5, 10 sec.), Battery Level

Indicator (5 Levels),

Unit selection (Air Velocity: m/s or FPM, Flow Rate: m

3

/min, m

3

/h,

ft

3

/min, or ft

3

/h, Temperature

*1

:

o

C or

o

F, Humidity

*2

: %RH, Static

Pressure

*3

: kPa or Pa)

Max., Min., and Average Calculation, Sampling Time: 1 to 999 sec.

No. of trails: 1 to 9999, Store up to 25 Duct Sizes (Square/Round, Size

Range: 1 to 9999mm or 0.1 to 999.9inch(diameter, height, width))

Max.Data Storage: 20000

Output

Digital Output: USB (when connected to a printer, automatically switch

to RS-232C, Baud Rate 4800, 9600, 19200 and 38400 bps)

Analog Output

*3

: DC 0 to 1V (select one from Air Velocity,

Temperature

*1

, Humidity

*2

and Pressure

*3

)

Power

6 × AA Batteries , AC Adaptor

*4

: AC 100 to 240V (50/60Hz)

Battery Life

Approximately 10 continuous hours (at air velocity 5m/s, 20

o

C, with

alkaline batteries)

Operating Environment

Main Body: 5 to 40

o

C (41 to 104

o

F) with no visible condensation

Probe: -20 to 70

o

C (-4 to 158

o

F) with no visible condensation

Storage Environment

-10 to 50

o

C (14 to 122

o

F) with no visible condensation

Weight

Approximately 400g (batteries included)

Accessories

Carrying case, Operation manual, 6×AA Batteries, Probe cable (2m),

USB cable

Options

Spare probe, Analog output, Pressure sensor, Extension rod, Printer,

Printer cable, Software (for Windows), AC adaptor, Hands-free Case

*

1: Air temperature is only available on MODEL6531/6541/6542/6533/6543/6561

*

2: Humidity is only available on MODEL6531/6533

*

3: Pressure is only available on MODEL6501-B0/-C0 , Analog Output is only available on MODEL6501-A0/-C0

*

4: Optional.

Advertising