KANOMAX 6501 series Climomaster User Manual

Page 61

Advertising
background image

60 Trouble Shooting

Output check (1) - Print out

Problem

Possible cause(s) / Solution(s)

Reference

The printer doesn’t

operate.

The printer is not connected properly.

Confirm the printer connection.

Page 34

The Baud Rate might not be set properly.

Check both the instrument and printer

settings.

Page 46

Make sure to use the optional printer (Model

DPU-S245) and printer cable (Model
6000-31).

Page 34

Confirm the connection procedure.

Connect the printer to the printer port,

then turn on the instrument and printer.

Page 34

Cannot printout the

display

Display is not frozen.
 1. Press to hold the display.

2. Press to print out.

Page 34

Want to cancel printing

You can’t cancel print while it is operating.

Page 34

Output check (2) - Digital output

Problem

Possible cause(s) / Solution(s)

Reference

PC doesn’t load the
measuring data.

Confirm the USB connection, which is
different from printer connection.

Page 39

The Baud Rate might not be set properly.

Check both the instrument and PC

settings.

Page 34

Confirm the communication command (ex.

Hyper Terminal command).

Page 39


Output check (3) - Analog output

Problem

Possible cause(s) / Solution(s)

Reference

No data is generated.

Check the connection and confirm the

polarity.

Page 43

The HOLD function might be active.

Press to cancel the HOLD function.

Page 43

Data is tiered.

Data is outputted every one second.

Page 43

The output data is not
correct.

Confirm the setting for analog output.

Page 43

Confirm the output range.

Page 43

The load impedance might not be optimum.
Set the load impedance larger than 5k

Ω.

Page 43


Advertising