Analog output, 9 analog output – KANOMAX 6501 series Climomaster User Manual

Page 44

Advertising
background image

43

§ 9

Analog Output

1.

Data Update……every 0.1 sec

(Except Humidity; updated every 1sec)

2.

Load Impedance……Above 5KΩ

3.

Output Voltage……DC 0 to 1V

For analog output, you must select one setting
from the table below.
(There is no analog output for Flow Rate)

Parameter

Output Range

Conversion Formula

(V: voltage)

Velocity

(U)

0 to 5 m/s

U= 5×V m/s

0 to 10 m/s

U= 10×V m/s

0 to 30 m/s

U= 30×V m/s

0 to 50 m/s

U= 50x V m/s

0 to 1000 FPM

U= 1000×V FPM

0 to 2000 FPM

U= 2000×V FPM

0 to 6000 FPM

U= 6000× V FPM

0 to 10000 FPM

U= 10000×V FPM

Humidity

(H)

0 to 50 %RH

H= 50×V %RH

0 to 100 %RH

H= 100×V %RH

Temperature

(T)

-20 to 30 °C

T= 50×V-20 °C

0 to 50 °C

T= 50×V °C

0 to 100 °C

T= 100×V °C

-4 to 86

o

F

T= 90×V - 4

o

F

32 to 122

o

F

T= 90×V+32

o

F

32 to 212

o

F

T= 180×V+32

o

F

Pressure

(P)

-2 to +2 kPa

P= 4×V- 2 kPa

-5 to +5 kPa

P= 10×V- 5 kPa

For the output range, the low end will be set at 0V and the high end
will be set at 1V. The voltage is linear.
The Time Constant will be in effect.

Mode

Ways To Take In Measured Data (Analog Output)

Explanation

TC1



Take 10

measurements each

second and indicate

the average as an

instantaneous value

at every 1 sec.

TC5

Output the average

measured value for

5 seconds at every

0.1 second. Data

shifts by 0.1 second.

TC10

Output the average

measured value for

10 seconds at every

0.1 second. Data

shifts by 0.1 second.

Analog Output

Terminal

0 5 10 15 20sec

Measuring time

0 0.5 1.0 1.5 2.0sec

0 2.5 5.0 7.5 10sec

Average of 10sec

Average of 5sec

Average of 1 sec

Analog output is only
available on
MODEL6501-A0/-C0.

To change the setting
of Time Constant,
refer to “To change
Time Constant” on
P.20.