MULTI FLOW RATE Mode

MULTI FLOW RATE is very similar to the SINGLE FLOW RATE MODE,
except it adds an additional parameter for the number of locations.
This allows you to take several air flow measurements (i.e.
traverses) and get a total average, for example measuring flow rate
from multiple ducts to get an overall average.

The value of each point is the
average of TRIAL(1) to TRIAL(N).
The value of each location is the
average of POINT(1) to
POINT(M). The value of each
location will be stored in the
memory.

Press

to enter the Main Menu.

Use

to select “4. FLOW RATE”

and press

.

Use

to select

“2. MULTI FLOW RATE” and press

.

As the same as the SINGLE mode, set
the following: 1. Sampling Time, 2. No.
Trial, 4. Data Storage, and 5. Duct
Entry No.

Use

to select

“3. MEAS./LOC.” and press

.

Use

to select the number of

MEAS. points (1 to 999) and press

.

Then use

to select the

number MEAS. locations (1 to 999) and

press

.

MULTI

(

R20000/20000

1.SAMPLING(TIME)001

p

s

2.SAMPLING(n)

11

0050

3.MEAS./LOC.

oo

010/01

0

4.DATA

2

STORAGE

2

?

22

YES

5.DUCT

2

ENTRY

2

NO.

222

1

6.SET

2

TO

2

START

SAMPLING MODE

1.SINGLE

o

FLOW

o

RATE

2

.MULTI

o

FLOW

o

RATE

LOCATION(1) :AVG(1)=ΣPOINT(i)/M
LOCATION(2) :AVG(2)=ΣPOINT(i)/M
・・・
LOCATION(L) :AVG(L)=ΣPOINT(i)/M

<Result>

Average(AVG): AVG=(AVG(1)+ AVG(2)+ ・・・・+AVG(L))/L

Maximum(MAX): MAX=Maximum value from AVG(1) to AVG(L)

Minimum(MIN): MIN=Minimum value from AVG(1) to AVG(L)

POINT(1) :avg(1)=ΣTRIAL(i)/N
POINT(2) :avg(2)=ΣTRIAL(i)/N
・・・
POINT(M)

:avg(M)=ΣTRIAL(i)/N