§ 15

Trouble Shooting

Check battery

Problem

Possible cause(s) / Solution(s)

Reference

The instrument will not

turn on

Battery is inserted in wrong polarity.

Turn off the power and insert the battery

correctly.

Page 11 &

13

The display is not visible

Display contrast is not set up properly.

Adjust the contrast volume switch.

Page 50

The battery level indicator

flashes on the display

The battery is low.

Turn off the instrument and replace the

batteries.

Page 11 &

13


Initial operational check

Problem

Possible cause(s) / Solution(s)

Reference

The display is not visible

The display contrast is not set up properly.

Adjust the contrast dial.

Page 50

[NO PROBE] is displayed

The probe is not connected.

Turn off the instrument and connect a

probe.

Page 12

Can I change the
measurement unit?

Yes, you can change measurement units by

following the steps on page 46.

Page 46


Normal operational check

Problem

Possible cause(s) / Solution(s)

Reference

[**.*] is displayed as

measured value.

It indicates the measurement is outside the
measuring range.

The instrument must be used in the range

stated in the specifications.

Page 52

[- - - - ] is displayed as air
velocity measurement.

The probe or the probe cable is not connected

properly.

Check the connection between the probe

and the probe cable.

Page 12

The probe or the probe cable may be

damaged.

Contact your local Kanomax

representative for repair or replacement.

Page 62

The instrument is not

reading the correct air
speed.

The probe might be not facing into the wind.

Make sure the wind direction indicator is

facing into the wind.

Page 14

The instrument reads a

higher temperature with
no wind.

CLIMOMASTER can’t measure the correct
temperature with no wind (less than 0.1m/s

of wind). *Model 6533 and 6543 can

measure less than 0.1 m/s.

Page 14

The response time is slow Set the correct time constant (TC).

Page 20

[ - - - - ] is displayed as

the air volumetric flow

measurement.

The duct information (shape and size) is not

stored.

Register duct information.

Page 16