40 Digital Output

Transmission of On-Time Data

(

Measured at every 1sec

)

<Enter the Number of Data Needed>
1. Type “D**** ”.

*

Type in the number of data(4 digits) after “D”.

2. After “AD”, the data will be displayed.
3. Each data represents 1 second of measurement. If you ask for 20

data, it takes approximately 20 seconds to display.

4. The maximum number that can be set is 9999. If you need

more, re-send the command


Output Content

Air Velocity Mode: Velocity; Temperature; Humidity

Humidity Flow Rate Mode: Flow Rate; Temperature; Humidity

Pressure Mode: 0000000; 0000000; Pressure


<To Cancel>

Type “N ”.


<To Download the Measuring Conditions>
1. Type “S ”
2. After “AS”, the data will be displayed as described below.

Output Content
Measuring Parameter (WTH, VTH or PRS); Time Constant (P.20);
Duct Size (Width, Height, Diameter); Units of the Duct Size

WTH: Flow Rate, Temperature, Humidity

VTH: Air Velocity, Temperature, Humidity

PRS: Pressure




<To Download the Measuring Units>
1. Type “U ”

2. After “AU”, the data will be displayed.

Output Content
Velocity Unit; Temperature Unit; Humidity Unit; Pressure Unit; Flow
Rate Unit

”0000000” will be
displayed for
humidity on
MODEL6541/
6542/6543/6561.

”0000000” will be
displayed for
temperature and
humidity on
MODEL6551 and
6552