Warning – Canon TX-20 Full Auto Tonometer User Manual

Page 83

Advertising
background image

7Troubleshooting

83

Warning

Code

Cause

Remedy

231

Auto-alignment stopped because the limit of motion (to
right) of the measurement unit was detected during auto-
alignment.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Be sure that the patient’s forehead and chin are firmly

against the forehead rest and chin rest and then do
measurement again.

232

Auto-alignment stopped because the limit of motion (to
left) of the measurement unit was detected during auto-
alignment.

233

Auto-alignment stopped because the limit of motion (up) of
the measurement unit was detected during auto-alignment.

234

Auto-alignment stopped because the limit of motion
(down) of the measurement unit was detected during auto-
alignment.

235

Auto-alignment stopped because the limit of motion
(forward) of the measurement unit was detected during
auto-alignment.

236

Auto-alignment stopped because the limit of motion
(backward) of the measurement unit was detected during
auto-alignment.

237

Auto-alignment stopped because the set stop position of
the nozzle was detected during auto-alignment.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Do the following and do the measurement again.

Press the LIMITER button to reset the position where the

nozzle stops.

Be sure that the patient’s head and chin are firmly

against the forehead rest.

238

Operation stopped because a button or the operation lever
was operated during auto-alignment.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Do the measurement again.

241

Measurement stopped because an intraocular pressure
measurement error occurred during full auto-alignment
measurement.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Do the measurement again.

242

The measured intraocular pressure is over the value set for
the [Upper IOP Warning].

243

The measured intraocular pressure is below the value set
for the [Lower IOP Warning].

244

Auto-alignment stopped because the measurement unit
and the chin rest may be interfering with each other.

1) Record the message code, contents (including error

codes) and then execute [OK].

2) Be sure that the patient’s head and chin are firmly against

the forehead rest.