Age1439_input_offset – Agilent Technologies VXI E1439 User Manual

Page 145

Advertising
background image

135

Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_input_offset

Sets the dc offset DAC setting for the current range. This description also includes the query:

age1439_input_offset_get

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_input_offset(ViSession id, ViInt16 coarseDac, ViInt16 fineDac);
ViStatus age1439_input_offset_get(ViSession
id, ViPInt16 coarseDacPtr, ViPInt16

fineDacPtr);

Description

These values are normally set by

age1439_input_autozero

so you generally would use this

command only for special situations. The resultant values can be saved to non-volatile RAM with

age1439_input_offset_save

.

Each ac coupling range has a unique DAC setting. All dc coupling ranges use the same DAC
setting as the highest range setting for ac coupling. The scaling between the coarse and fine DACs
is approximately 100 to 1.

AGE1439_OFFS_DAC_MIN

sets the minimum dc offset DAC setting.

AGE1439_OFFS_DAC_MAX

sets the maximum dc offset DAC setting.

Parameters

id

is the VXI instrument session pointer returned by the age1439_init function.

coarseDac

sets values of 0 to 255.

fineDac

sets values of 0 to 255.

coarseDacPtr

returns a pointer to the current value of coarseDac

fineDacPtr

returns a pointer to the current value of fineDac

Return Value

AGE1439_SUCCESS indicates that a function was successful.

Values other than AGE1439_SUCCESS indicate an error condition or other important status
condition. To determine the error message, pass the return value to

“age1439_error_message” on

page 102

.

See Also

“Default values” on page 201

,

“age1439_init” on page 132

,

“age1439_input_autozero” on

page 134

,

“age1439_input_offset_save” on page 136