Age1439_lbus_reset – Agilent Technologies VXI E1439 User Manual

Page 160

Advertising
background image

150

Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_lbus_reset

Resets the local bus (Agilent E1439D only) . This description also includes the query:

age1439_lbus_reset_get gets the current local bus reset state

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_lbus_reset(ViSession id, ViInt16 lbusReset);
ViStatus age1439_lbus_reset_get(ViSession
id, ViPInt16 lbusResetPtr);

Description

In order to avoid glitches in the local bus data, the local bus interface has strict requirements as to
the order in which modules in a VXI mainframe have their local bus interface reset. Upon power-
up or whenever any single module in the mainframe is put into a reset state, all modules should be
placed into the reset state from left to right. Then all modules can be take out of reset from left to
right.

Parameters

id

is the VXI instrument session pointer returned by the age1439_init function.

lbusReset

puts the Agilent E1439's local bus into reset or takes it out of reset.

AGE1439_LBUS_RESET_ON

puts the Agilent E1439's local bus into reset while

AGE1439_

LBUS_RESET_OFF

takes the Agilent E1439 out of reset.

lbusResetPtr

points to the current value of the lbusReset parameter.

Return Value

AGE1439_SUCCESS indicates that a function was successful.

Values other than AGE1439_SUCCESS indicate an error condition or other important status
condition. To determine the error message, pass the return value to

“age1439_error_message” on

page 102

.

See Also

“Default values” on page 201

,