44

Using the Agilent E1439

Fiber Optic Interface

Fiber Frames

Data is transmitted over the fiber interface in a series of fiber frames. Each fiber frame is
composed of a series of 32-bit values, which encode to 40 bits. Each 32-bit value can either be
data or an ordered set. Data and ordered sets are strung together to make the three types of fiber
frames—Data Frame, BOF, and EOE. The Data Frame transmits 0 to 512 32-bit data words. The
exact amount of data that is sent depends on the amount of data that is available when the fiber
interface is ready to send the Data Frame. BOF (Beginning Of Frame) is a synchronizing event
that can be sent just prior to the start of data transmission. EOE (End Of Epoch) is a synchronizing
event that contains the last 4 data bytes in an epoch. An epoch is composed of one or more Data
Frames followed by an EOE. The following shows the ordered sets and data that make up the
three fiber frames:

Control Signals

PIO1, PIO2, DIR, and NRDY are FPDP (front panel data port) control signals. These signals can
be defined by another product or you can define their meaning and application.

When an overflow condition in the transmit FIFO occurs, the E1439D asserts FIFO OV indicating
a loss of data. This may occur in

Append

fiber mode if the available fiber bandwidth capability is

exceeded.

If flow control is enabled, the E1439D responds to the STOP and GO signals. See

“Generate” on

page 48

for details.

Data Frame (Normal Data Fiber Frame)

IDLE

1

1. Pad for Data Frame or BOF

SOF

2

2. Start Of Frame, framing event that embeds PIO1, PIO2, and DIR

0 to 512 data words

3

3. 32 bit or 4 Byte words, maximum 2048 Bytes

CRC

4

4. Cyclic Redundancy Check, optional

FEOF

5

5. Frame End Of Frame, end of Data Frame

SEOF

6

6. Status End Of Frame, embeds FIFO OV and NRDY

GO/STOP

7

7. Flow controls

BOF (Sync Without Data Fiber Frame)

IDLE

1

SOF

2

No data

CRC

4

MEOF

8

8. Mark End Of Frame, end of BOF and EOE

SEOF

6

GO/STOP

7

EOE (Sync with Data Fiber Frame)

SWDV

9

9. Sync With Data Valid, start of EOE

SOF

2

Last 4 data bytes in epoch

10

10.4 bytes, exactly

CRC

4

MEOF

8

SEOF

6

GO/STOP

7