age1439_status_get

Reads status register information for the module.

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_status_get(ViSession id, ViPInt16 statusPtr);

Parameters

id

is the VXI instrument session pointer returned by the age1439_init function.

statusPtr

points to the status word. The bits are defined below:

Status Bit

Definition

Description

0-1

AGE1439_STATUS_MEAS_ARM_WAIT
AGE1439_STATUS_MEAS_IDLE
AGE1439_STATUS_MEAS_IN_
PROGRESS
AGE1439_STATUS_MEAS_TRIG_WAIT

These two bits indicate the current state of the
measurement loop. See

“The measurement loop” in

chapter 3

and

“age1439_meas_status_get” on page 156

for more information about these states

2

AGE1439_STATUS_PASSED

Passed: This bit is always set to 1

3

AGE1439_STATUS_READY

This bit is set when the module is ready after power-on. See
the VXIbus Specifications for more information.

4

AGE1439_STATUS_FIBER_ACTIVE

This bit is set internally whenever any data has been written
to the receive FIFO, or read from the transmit FIFO of the
fiber interface within the past 500 milliseconds,
approximately. The bit is cleared automatically when activity
ceases on the fiber interface

5

AGE1439_STATUS_FIBER_ERROR

This bit is set internally whenever an error condition occurs
on the fiber interface. Reading the status register does not
clear this bit. To do this, the

age1439_fiber_error_clear

function must be used explicitly. The function

age1439_

fiber_error_get

can be used to read the contents of the fiber

error register. If the error is continuously present, the bit will
not be cleared.

6

AGE1439_STATUS_SETUP_ERROR

Setup error: An invalid parameter value was requested. If an
invalid block size was requested, the closest valid block size
is used until a change to an interrelated parameter makes the
requested block size valid. If a data resolution, data type,
filter bandwidth, trigger delay, or filter decimation parameter
was requested which would result in an inability to make a
measurement, the previous valid parameter is used until a
change to an interrelated parameter makes the requested
parameter valid

7

AGE1439_STATUS_SYNC_COMPLETE

Sync/Idle Complete: This bit is set when the most recent
user-initiated Sync or Idle change has propagated through to
all modules in a system. The change is a result of asserting
Sync or forcing Idle via the Control Register or issuing a
meas_control command or function