47

Using the Agilent E1439

Fiber Optic Interface

Raw

The Raw fiber mode transmits raw (i.e., unprocessed, full bandwidth) ADC data over the fiber
interface. At the same time that the raw data is transmitted over the fiber interface, filtered ADC
data can be sent over the local bus (Agilent E1439D only) or read from the FIFO via the VME
bus. After selecting Raw, optical data transmission starts at the trigger event and is not affected by
trigger delays or data delays. The raw data transmission continues even after the measurement is
complete. Changing the fiber mode stops data transmission. See the following illustration.

Note

Setting the data port to Fiber while in the Raw fiber mode results in an invalid instrument state
because raw and filtered data cannot both be sent over the fiber interface at the same time.

Fiber Interface Setup

Fiber Mode

Raw

Rate

106

1

or 250 MBs

1. Only with external sam-

ple. Internal sample gen-
erates data too fast for
this rate.

BOF

Optional

CRC

ON

2

2. Some receivers may

require CRC to be off
for compatibility

Flow Control

Optional

Epoch Generate

Optional

Epoch Size

Divisible by 4

Fiber TX

Fiber RX

VME BUS

LBUS

ADC

DIGITAL

FILTERS

FIFO

E1438D / E1439D