age1439_self_test

Performs a self-test and returns the result of that self test.

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_self_test(ViSession id, ViPInt16 testResult, ViChar testMessage[]);

Description

The Agilent E1439 self test includes the following tests:

Digital: verifies the integrity of paths from LO chip through the filters to the memory
controller.

Serial: verifies the integrity of serial setup path for each board.

Memory: fills the entire DRAM then verifies that all the data is correct.

Analog: verifies that auto zero adjust is working and that the input is triggering.

Clock: verifies that the oscillator is working properly.

Fiber: performs five-second internal fiber verification.

Parameters

id

is the VXI instrument session pointer returned by the age1439_init function.

testMessage

points to the self test status message string up to 256 characters long.

Note

For this parameter you must allocate a character array of at least 256 characters

AGE1439_STR_

LEN_MIN

, including the null byte, prior to calling this function in any programming language.

testResult

points to the instrument numeric error code.

Possible test result values are:

Error

Message

Error Code

(hex)

Self Test

Status Message

AGE1439_ST_SUCCESS

0x000

self test successful

AGE1439_ST_HARDWARE_FAIL

0x001

hardware failure

AGE1439_ST_SERIAL1_FAIL

0x002

serial 1 test failed

AGE1439_ST_SERIAL2_FAIL

0x004

serial 2 test failed

AGE1439_ST_CLOCK_FAIL

0X008

95 MHz clock test failed

AGE1439_ST_MEMORY_FAIL

0x020

memory test failed

AGE1439_ST_DIGITAL1_FAIL

0x040

real data path failed

AGE1439_ST_DIGITAL2_FAIL

0x080

complex data path failed