Front panel description, Module description – Agilent Technologies VXI E1439 User Manual

Page 218

Advertising
background image

208

Module Description

Front Panel Description

Front Panel Description

Intermodule

(E

CL

)

Clock

Sync

Ext Clock/Ref

Ext Trigger

Analog In

5Vrms Max

Access

Overload

LED lights when the module is accessed

via the VXI backplane.

Sync extenders are used to extend the

sync line from one mainframe

or module to another. It is an SMB connector

for ECL levels and must be terminated in

50 ohms at each end of the chain.

BNC input for ECL or TTL signals that can trigger

data acquisition. For ECL, the input is ac coupled,

1 k ohm, edge sensitive. For TTL, the input is

dc coupled, 1 k ohm, TTL levels.

This is the main input to the ADC. It is a

single-ended input terminated into 50 ohms.

Clock Extenders are used to connect the

system reference from one mainframe or

module to another. It is an SMB connector

for ECL levels and must be terminated in

50 ohms at each end of the chain.

BNC input for TTL, ECL, or sine

wave signals that can be used as the

ADC sample clock. This input can also

be used for the system frequency reference.

This input is ac coupled, and has 1 k ohm

impedance.

XMT RCV

I/O

Data

XMT

RCV

LED lights when the transmitter is enabled.

LED blinks when data generated by this module

is being transmitted. (”D” module only)

LED lights when an optical signal is detected.

LED blinks when data is being received.

(”D” module only)

Fiber optic serial FPDP data link.

Dual LC connector.

(”D” module only)

exceeded, producing an overload in the ADC

LED lights whenever the input range is

Advertising