Using the agilent e1439 – Agilent Technologies VXI E1439 User Manual

Page 56

Advertising
background image

46

Using the Agilent E1439

Fiber Optic Interface

Copy

The Copy fiber mode copies optical data from its fiber receiver to its fiber transmitter without
adding any data. Normal data collection and filtering continues, and the data port selection
determines whether data is sent to the local bus (Agilent E1439D only) or read from the FIFO via
the VME bus. Copy is the default fiber mode after power-on or reset. See the following
illustration.

Note

Setting the data port to Fiber while in the Copy fiber mode results in an invalid instrument state.

Fiber Interface Setup

Fiber Mode

Copy

Rate

106 or 250 MBs

BOF

setting ignored

CRC

must match incoming signal

Flow Control

setting ignored

Epoch Generate

setting ignored

Epoch Size

setting ignored

Fiber TX

Fiber RX

VME BUS

LBUS

ADC

DIGITAL

FILTERS

FIFO

E1438D / E1439D

FIFO

1 KB