The pipetting programs, Pipmix, Revpip – BrandTech Scientific Transferpette electronic User Manual

Page 45: Disp

Advertising
background image

English

GEL mode is not available for Transferpette

®

electronic

1000 µl and 5000 µl.

The Pipetting Programs

1.

Normal Pipetting

PIP

Mode

Standard program. A previously set volume
is aspirated into the pipette tip and then
discharged.

46

4.

Pipetting for Electrophoresis

GEL

Mode

Program for loading electrophoresis gels. A
predefined sample volume is aspirated at high,
adjustable speed and then slowly discharged.

52

2.

Mixing of Samples

PIPmix

Mode

Program for mixing liquids. The sample is
repeatedly aspirated and discharged.

48

3.

Reverse Pipetting

revPIP

Mode

Program especially for pipetting liquids with a
high viscosity or vapor pressure, or that tend
to foam.

50

5.

Dispensing

DISP

Mode

Program for dispensing liquids. An aspirated
volume is dispensed repeatedly in defined
steps.

54