Electrophoresis (gel) mode – BrandTech Scientific Transferpette electronic User Manual

Page 52

Advertising
background image

1x

1x

1x

3x

Press the MENU key three
times to bring up the mode
selection menu.
‚Mode‘ blinks.

Scroll through the modes
using the arrow keys until
‚GEL‘ appears.
‚Mode‘ continues to blink.

Press the ENTER key.
The Display now shows ‚blo‘
for blow-out.

Press the pipetting key
once. The set volume is
aspirated into the tip.

1.

Bring up the menu

2.

Select GEL mode

3.

Confirm GEL mode

5.

Aspirate liquid

Press the pipetting key once
to move the piston into the
start position.
The arrow in the display
points upwards (aspiration).

4.

Prepare for

pipetting

Aspirate a larger volume
In order to aspirate a larger
volume than was set (up
to a max. of 110% of the
nominal volume), press
and hold the pipetting key
until the desired volume
has been aspirated. The
display shows a rhombus.

Electrophoresis (GEL) Mode

The Programs · Settings and Use

Program for loading electrophoresis gels. A predefined sample volume is aspirated into
the pipette tip with high adjustable speed and then slowly discharged.
Volume and speed adjustment is described on pages 42 and 43.

What to do

How to do it

Keys to press

Display readout