Servicing and cleaning, Gh i j g' j – BrandTech Scientific Transferpette electronic User Manual

Page 62

Advertising
background image

G

H

I

J

G'

J

Servicing and Cleaning

1.

Press both snap-in locks (J ) simultaneously

and remove lower part of ejector (G‘).

2.

Unscrew and remove the pipette shaft (H+I )

from the hand grip.

3.

Separate the magnetic connection between

both components by gently but firmly pulling

in opposite directions and remove upper part

of ejector (G).

4.

Unscrew piston unit (H) from lower part of

the pipette shaft ( I ).

5.

Remove the O-ring from the piston unit and

clean it.

6.

Clean piston unit (H) and lower part of pipette

shaft ( I ) with a mild soap solution or isopropanol

and then rinse with distilled water.

7.

Allow the parts to dry (max. 120 °C/248 °F)

and to cool down.

8.

Carefully lubricate O-ring inside and outside

and put it back in place.

9.

Assemble the cooled parts in reverse order

from above.

10.

Perform reference run (rEF).

Transferpette

®

electronic 500-5000 µl

In order to assure proper funtioning, the Transferpette

®

electronic should be serviced and cleaned at regular
intervals.

Inspect the pipette tip cone for damage.
Inspect the piston and seal for contamination and damage.
Test the sealing of the instrument. We recommend using
the BRAND leak testing instrument PLT unit. Alterna-
tively: to do this aspirate a sample, and then hold the
instrument in a vertical position for about 10 sec. If a drop
forms at the tip orifice, see the troubleshooting guide on
page 65.

Servicing

Disassembly and cleaning

Do not disassemble piston unit
(H) any further!

Note:

Individual components shown can be ordered as spare parts.
For ordering information see page 67.