Revpip mode – BrandTech Scientific Transferpette electronic User Manual

Page 50

Advertising
background image

3x

1x

1x

1x

1x

Scroll through the modes
using the arrow keys until
‚revPIP‘ appears.
‚Mode‘ continues to blink.

Press the pipetting key
once. The volume aspirated
will be a little bit more than
set.

To discharge the measured
amount of liquid, press the
pipetting key once. The
arrow in the display points
downwards (discharge).
Some liquid will remain
in the tip.

revPIP Mode

The Programs · Settings and Use

Press the MENU key three
times to bring up the mode
selection menu.
‚Mode‘ blinks.

1.

Bring up the menu

2.

Select revPIP

mode

Press the ENTER key.
The Display now shows ‚blo‘
for blow-out.

Press the pipetting key once
to move the piston into the
start position.
The arrow in the display
points upwards (aspiration).

3.

Confirm revPIP

mode

4.

Prepare for

pipetting

5.

Aspirate liquid

6.

Discharge liquid

in the revPIP

mode

Program for pipetting of liquids with high viscosity, vapor pressure or that tend to foam.
Volume and speed adjustments are described on pages 42 and 43.

What to do

How to do it

Keys to press

Display readout

50