Easy calibration – BrandTech Scientific Transferpette electronic User Manual

Page 59

Advertising
background image

1x

1x

>3s

English

Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
‚on‘. ‚CAL‘ blinks.

Press one of the arrow keys
to deactivate the CAL mode.
The display changes from
‚on‘ to ‚off‘. ‚CAL‘ continues
to blink.

Press the ENTER key.
The CAL symbol disap-
pears. The instrument has
now been reverted to factory
default setting.

The continually displayed CAL symbol refers to a previously made
adjustment.

1.

Bring up the CAL

mode

2.

Deactivate CAL

mode

3.

Revert to factory

setting

The instrument is permanently adjusted for watery solutions,
but it can also be set for solutions with varying density, vis-
cosity and temperature. The Transferpette

®

electronic can be

adjusted in every mode, with the exception of the GEL mode.

When the Transferpette

®

electronic is adjusted, a volume

offset is performed, which means that the volume is changed
across the entire volume range of the pipette by the same
amount. It is recommend that the adjustment be performed
at 50% of the nominal volume.

Easy Calibration

What to do

How to do it

Keys to press

Display readout

Revert to factory default settings

Important:

59

Note: