Pip mode – BrandTech Scientific Transferpette electronic User Manual

Page 46

Advertising
background image

3x

1x

1x

1x

Press the MENU key three
times to bring up the mode
selection menu. ‚Mode‘
blinks.

Use one of the arrow keys
to scroll through the modes
until ‚PIP‘ appears.
‚Mode‘ continues to blink.

Press the ENTER key.
The display now shows ‚blo‘
for blow-out.

Press the pipetting key once
to aspirate the liquid.

Press the pipetting key once
to move the piston into the
start position.
The arrow in the display
points upwards (aspiration).

PIP Mode

The Programs · Settings and Use

The standard program – a previously set volume is aspirated and then discharged.
Volume and speed adjustments are described on pages 42 and 43.

1.

Bring up the menu

2.

Select PIP mode

3.

Confirm PIP mode

4.

Prepare for

pipetting

5.

Aspirate liquid

What to do

How to do it

Keys to press

Display readout

46