Disp mode – BrandTech Scientific Transferpette electronic User Manual

Page 54

Advertising
background image

3x

1x

1x

+

1x

Scroll through the modes
using the arrow keys until
‚DISP‘ appears.
‚Mode‘ continues to blink.

Press the arrow keys (+/-)
to set the volume. Holding
the arrow key down accel-
lerates the rate of change.
‚VOL‘ blinks.

Press the ENTER key.
The display now shows the
new volume setting for the
dispensing steps and the
max. number of steps.
‚Steps‘ blinks.

DISP Mode

The Programs · Settings and Use

Press the MENU key three
times to bring ap the mode
selection menu.
‚Mode‘ blinks.

1.

Bring up the menu

2.

Select DISP

mode

Press the ENTER key.
The Display now shows
‚blo‘ for blow-out.

Press the pipetting key once
to move the piston into the
start position.
The arrow in the display
points upwards (aspiration).

3.

Confirm DISP

mode

4.

Prepare for

dispensing

5.

Set dispensing

step volume

6.

Confirm dispens-

ing step volume

Program for discharging an aspirated liquid in pre-defined steps.
The volume aspirated will be a little bit more than actually needed.
Speed adjustment is described on page 43.

What to do

How to do it

Keys to press

Display readout

54

Advertising