ISO 9001 and GLP-guidelines require regular examinations of your volumetric
instruments. We recommend checking the volume every 3-12 months. The
interval depends on the specific requirements on the instrument. For instru-
ments frequently used or in use with aggressive media, the interval should be
shorter. The detailed testing instruction can be downloaded on www.brand.de
or www.brandtech.com.
BRAND also offers you the possibility to have your instruments calibrated
by the BRAND Calibration Service or the BRAND-owned DAkkS Calibration
Service.
Just send in the instruments to be calibrated, accompanied by an indication
of which kind of calibration you wish. Your instruments will be returned within
a few days together with a test report (BRAND Calibration Service) or with
a DAkkS Calibration Certificate. For further information, please contact your
dealer or BRAND. Complete ordering information is available for download at
www.brand.de (see Technical Documentation).

Calibration Service

We shall not be liable for the consequences of improper handling, use, servic-
ing, operating or unauthorized repairs of the instrument or the consequences
of normal wear and tear especially of wearing parts such as pistons, seals,
valves and the breakage of glass as well as the failure to follow the instruc-
tions of the operating manual. We are not liable for damage resulting from
any actions not described in the operating manual or if non-original spare
parts or components have been used.