C

B

A

English

c) Put the battery compartment cover in
place again and close it.

Replace the battery

a) Open the battery compartment cover.
Remove the battery and pull the plug
gently out of the socket.

b) Insert the plug of the new battery into the
socket and insert the battery.

Charge the battery

A fully charged battery allows approx. eight hours (equals more than 4000 pipetting cycles) of
continuous pipetting of liquids with a viscosity and density similar to water.

a) Insert the charging cable plug for the

AC adapter into the jack at the top of the
Transferpette

®

electronic; charging starts

automatically

b) During the charging, the bars for the battery
capacity run continually from the bottom to
the top. The battery is fully charged, when
the bars in the display have stopped
moving.

Before charging the battery ensure that the AC adapter is compatible with the line voltage in the labora-
tory. Do not charge the device in an explosive environment. The battery can only be charged inside the
Transferpette

®

electronic.

Remove the battery from the instrument, when it is not to be used
for longer periods.

Charging and Replacing the Battery

Important!

During charging, you can continue to work with the Transferpette

®

electronic.
If the battery is fully discharged, it will take a few minutes until a certain
minimum charge capacity is available, which is needed to operate the
instrument safely.

Pipetting during charging?

The last settings are stored in the memory of the instrument. If the battery is fully
discharged or the battery is changed, these settings are saved.

Note: