Operating Instructions Temperature Indicating Controller CF2S

V1.2

08/2004 - 9 -

Select the sensor type by rotary switch sensor input (SW303).
Factory adjusted as K, [ ].

[Table 3.2-2]

Rotary Sw.

No.

DIP Sw.
No. 7

Type of

sensor

Scale range

0

OFF

K

-200 to 1370 °C

-320 to 2500 °F

1

OFF

J

-200 to 1000 °C

-320 to 1800 °F

2

OFF

R

0 to 1760 °C

0 to 3200 °F

3

OFF

B

0 to 1820 °C

0 to 3300 °F

4

OFF

PL-

0 to 1390 °C

0 to 2500 °F

5

OFF

N

0 to 1300 °C

0 to 2300 °F

6

OFF

Pt100

-199.9 to 850.0 °C

-199.9 to 999.9 °F

7

OFF

JPt100

-199.9 to 500.0 °C

-199.9 to 900.0 °F

0

ON

S

0 to 1760 °C

0 to 3200 °F

1

ON

E

0 to 1000 °C

0 to 1800 °F

2

ON

T

-199.9 to 400.0 °C

-199.9 to 750.0 °F

3

ON

C (W/Re5-26)

0 to 2315 °C

0 to 4200 °F

7

ON

Pt100

-200 to 850 °C

-320 to 1560 °F

Advertising