28.4.4. Set tare value

Syntax: UT_TARE CR LF, where TARE – tare value

Possible answers:

UT_OK CR LF

- command executed

UT_I CR LF

- command comprehended but cannot be executed

ES CR LF

- command not recognised (possible wrong tare format)


Notice:
This protocole uses the dot character as a decimal point for separating
the decimal fraction part.



28.4.5. Send the stable result in basic unit

Syntax: S CR LF

Possible answers:


S_A CR LF
S_E CR LF

- command accepted and in progress
- time limit for stable result exceeded

S_I CR LF

- command comprehended but cannot be executed

S_A CR LF
MASS FRAME

- command accepted and in progress
- mass value in basic unit is returned


Frame format:

1 2-3

4

5 6 7-15 16 17

18

19

20

21

S space stability space sign

mass

space unit CR

LF


Example:

S CR LF – computer command
S _ A CR LF - command accepted and in progress

S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.