Autozero – RADWAG WPW/T 60/HR3/FH Waterproof Control Scales User Manual

Page 32

Advertising
background image

AUTO

- disable automatic tare (the mode is remembered after

restart);

NORMAL

-

tarring by pressing

;

MEMORY

- tare memory mode - the last tare value is being kept in

a non-volatile memory, Net pictogram is displayed.

SUM

- sum of tares – summing up a product tare value with

a tare from the database of tare values or with an
inscribed one.

Return to weighing:

See 11.3.



13.6. Autozero

The autozero function has been implemented in order to assure precise
indications. This function controls and corrects „0” indication. While the
function is active it compares the results continuously with constant
frequency. If two sequentional results differ less than the declared
value of autozero range, so the scale will be automatically zeroed
and the pictograms

and

will be displayed.

When AUTOZERO is disabled zero is not corrected automatically.
However, in particular cases, this function can disrupt the measurement
process e.g. slow pouring of liquid or powder on the weighing pan.
In this case, it is advisable to disable the autozero function.

Procedure:

• Enter

<P1 SCALE PARAMETERS> according to 11.2. and then:

- 32 -

Advertising