Send the result immediately in current unit, Switch on continuous transmission in basic unit, Witch on continuous transmission in basic unit – RADWAG WPW/T 60/HR3/FH Waterproof Control Scales User Manual
Page 146: Frame format
- 146 -
28.4.9. Send the result immediately in current unit
Syntax: SUI CR LF
Possible answers:
SUI_I CR LF
- command comprehended but cannot be executed
SUI_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in current unit is returned immediately
Frame format:
1 2 3
4
5
6 7-15 16 17 18 19 20 21
S U I stability space sign mass space
unit
CR LF
Example:
S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed and mass
returned
28.4.10. Switch on continuous transmission in basic unit
Syntax: C1 CR LF
Possible answers:
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1 2
3
4
5
6 7-15
16 17 18 19 20 21
S I space stability space sign mass space
unit
CR LF