18.3. Respond message format

After sending a request message you can receive:

XX_A CR LF

command accepted and in progress

XX_D CR LF

command completed (appears only after XX_A)

XX_I CR LF

command comprehended but cannot be executed

XX _ ^ CR LF

command comprehended but time overflow error appeared

XX _ v CR LF

command comprehended but the indication below the

XX _ OK CR LF Command done

ES_CR LF

Command not comprehended

XX _ E CR LF

error while executing command – time limit for stable result
exceeded (limit time is a descriptive parameter of the scale)


XX

- command name

_

- substitutes spaces

18.4. Command’s description

18.4.1. Zeroing

Syntax Z CR LF

Possible answers:

Z_A CR LF
Z_D CR LF

- command accepted and in progress

- command completed

Z_A CR LF
Z_^ CR LF

- command accepted and in progress

- command comprehended but zero range overflow appeared

Z_A CR LF
Z_E CR LF

- command accepted and in progress

- time limit for stable result exceeded

Z_I CR LF

- command comprehended but cannot be executed



18.4.2. Tarring

Syntax: T CR LF

Possible answers:

Advertising