Get tare value, Send the stable result in basic unit – RADWAG PUE C/31H/Z Indicator User Manual

Page 57

Advertising
background image

- 57 -

T_A CR LF
T_D CR LF

- command accepted and in progress
- command completed

T_A CR LF
T_v CR LF

- command accepted and in progress
- command comprehended but tare range overflow appeared

T_A CR LF
T_E CR LF

- command accepted and in progress

- time limit for stable result exceeded

T_I CR LF

- command comprehended but cannot be executed



18.4.3. Get tare value

Syntax: TO CR LF

Possible answers:

TO_TARA

CR LF

- command executed


Frame format:

1 2

3

4

5-6 7-15

16 17 18 19 20 21

T O space stability space tare space

unit

CR LF

Tare - 9 characters with decimal point justified to the right
Unit

- 3 characters justified to the left



18.4.4. Send the stable result in basic unit

Syntax: S CR LF

Possible answers:

S_A CR LF
S_E CR LF

- command accepted and in progress
- time limit for stable result exceeded

S_I CR LF

- command comprehended but cannot be executed

S_A CR LF
MASS FRAME

- command accepted and in progress
- mass value in basic unit is returned

Advertising