Send the result immediately in basic unit, Send the stable result in current unit, In basic unit – RADWAG PUE C/31H/Z Indicator User Manual

Page 58: Send the result immediately, Frame format, Syntax: si cr lf, Possible answers

Advertising
background image

- 58 -

Frame format:

1 2-3

4

5 6 7-15 16 17

18

19

20

21

S space stability space sign

mass

space unit CR

LF


Example:

S CR LF – computer command
S _ A CR LF - command accepted and in progress

S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done,
mass value in basic unit is returned.



18.4.5. Send the result immediately in basic unit

Syntax: SI CR LF

Possible answers:

SI_I CR LF

- command comprehended but cannot be executed at the
moment

MASS FRAME

- mass value in basic unit is returned


Frame format:

1 2

3

4

5

6 7-15

16 17 18 19 20 21

S I space stability space sign mass space

unit

CR LF

Example:

S I CR LF – computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass
value in basic unit is returned immediately.

18.4.6. Send the stable result in current unit

Syntax: SU CR LF

Possible answers:

Advertising
This manual is related to the following products: