Send the result immediately in current unit, In current unit, Send the result immediately – RADWAG PUE C/31H/Z Indicator User Manual

Page 59: Frame format

Advertising
background image

- 59 -

SU_A CR LF
SU_E CR LF

- command accepted and in progress
- timeout while waiting for stable results

SU_I CR LF

- command comprehended but cannot be executed

SU_A CR LF
MASS FRAME

- command accepted and in progress
- mass value in current unit is returned


Frame format:

1 2

3

4

5

6 7-15 16 17 18 19 20 21

S U space stability space sign mass

space

unit

CR

LF

Example:

S U CR LF – computer command
S U _ A CR LF - command accepted and in progress

S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass
value in current unit is returned.



18.4.7. Send the result immediately in current unit

Syntax: SUI CR LF

Possible answers:


SUI_I CR LF

- command comprehended but cannot be executed

MASS FRAME

- mass value in current unit is returned immediately


Frame format:

1 2 3

4

5

6 7-15 16 17 18 19 20 21

S U I stability space sign mass space

unit

CR LF


Example:

S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed
and mass returned

Advertising