LKPv and LGPv laboratory appliances with Profi-Controller

LKPv 8420

Gradient H0- :

1.74 K

(+0.9 K / – 0.84 K from

set temperature)

Max. fluctuation:

1.5 K

8,0 °C

7,5 °C

7,0 °C

6,5 °C

6,0 °C

5,5 °C

5,0 °C

4,5 °C

4,0 °C

3,5 °C

3,0 °C

2,5 °C

2,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

H0 = Gradient

Probes

Set temperature

Average probe temperature

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

8,0 °C

7,5 °C

7,0 °C

6,5 °C

6,0 °C

5,5 °C

5,0 °C

4,5 °C

4,0 °C

3,5 °C

3,0 °C

2,5 °C

2,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

LKPv 1420

H0 = Gradient

Gradient H0- :

2.0 K

(+0.5 K / – 1.5 K from

set temperature)

Max. fluctuation:

1.2 K

Probes

Set temperature

Average probe temperature

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty