Liebherr WKt 4552 GrandCru User Manual
Page 29
29
LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller
LGex 3410
t1
t2
t3
t4
t5
t6
t7
t8
t9
Probes
-17.0 °C
-17.5 °C
-18.0 °C
-18.5 °C
-19.0 °C
-19.5 °C
-20.0 °C
-20.5 °C
-21.0 °C
-21.5 °C
-22.0 °C
-23.0 °C
H0 = Gradient
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C
t1
t2
t3
t4
t5
t6
t7
t8
t9
Probes
LKexv 3910
H0 = Gradient
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty
Gradient H0- :
2.9 K
(+1.4 K / – 1.5 K from
set temperature)
Max. fluctuation:
0.40 K
Gradient H0- :
2.8 K
(+1.2 K / –1.6 K from
set temperature)
Max. fluctuation:
3.4 K
Set temperature
Average probe temperature
Set temperature
Average probe temperature