Liebherr WKt 4552 GrandCru User Manual
Page 27
27
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C
t1
t2
t3
t4
t5
t6
t7
t8
t9
Probes
LKUexv 1610
H0 = Gradient
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty
LGUex 1500
t1
t2
t3
t4
t5
t6
t7
t8
t9
Probes
-17.0 °C
-17.5 °C
-18.0 °C
-18.5 °C
-19.0 °C
-19.5 °C
-20.0 °C
-20.5 °C
-21.0 °C
-21.5 °C
-22.0 °C
-22.5 °C
-23.0 °C
-23.5 °C
-24.0 °C
H0 = Gradient
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty
Gradient H0- :
5.7 K
(+2.1 K / – 3.6 K from
set temperature)
Max. fluctuation:
0.80 K
Gradient H0- :
4.1 K
(+1.6 K / –2.5 K from
set temperature)
Max. fluctuation:
3.0 K
LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller
Set temperature
Average probe temperature
Set temperature
Average probe temperature