External alarm contact connection – Liebherr WKt 4552 GrandCru User Manual

Page 7

Advertising
background image

7

+

-

+

-

Normal operation

Alarm condition

e.g. green lamp

indicating normal

opretaion

external power supply

e.g. red lamp

indicating alarm

condition

external power supply

N.O.

COM

N.C.

N.O.

COM

N.C.

Alarm condition

Normal operation

Open or close A1– 6 to GND

Alarm inputs A1– 6

Any contractors

normally open or closed

GND

GND

GND

GND

A1

A2

A3

A4

A5

A6

COM

COM

N.O.

N.C.

External alarm contact connection

LKPv and LGPv laboratory appliances with Profi-Controller

Example: GSM dialler alarm system using volt free alarm contact

Alternatively, if no internal warning system is in place and additional wiring inside the building is problematical, the volt-free contact can be used in conjunction
with e.g. a GSM alarm dialler. There are many different versions of GSM alarm dialers available on the market which either call a telephone number and/or send
a text message in case of an alarm. Depending on the specification of the alarm dialler, potentially many appliances can be connected to one dialler. Being GSM
appliances, only a SIM card and a power supply are required.

The following example shows a circuit with a GSM dialler: A change in relay setting activates the alarm function of the dialler and thus initiates either a telephone
call or an SMS to pre-defined numbers with a pre-recorded message. The dialler in the example has 6 addressed alarm inputs. This could be 6 individual appliances
or 6 rooms, each with several appliances wired in series.

Volt-free contact
All LKPv and LGPv models have an integrated volt-free contact. The relay contact can be wired into a warning system in the building which would advise e.g. security
personnel or a janitor outside of working hours. Warning systems with audio or visual alarms are possible.

The following example shows a circuit with a visual alarm. Either normal operation or alarm condition or both can be indicated. Either AC or DC circuits are possible,
max AC 230 V / 5 Amp.

Example: visual alarm system using volt free alarm contact

In spite of all efforts to produce appliances of unequalled reliability, Liebherr strongly recommends that the volt-free contact is wired into some kind of independent
and external warning system. Please take a few moments to consider the value of the contents of the refrigerators and freezers under your supervision. The cost
of such a warning system is often negligible in comparison to the value of the contents.