Liebherr WKt 4552 GrandCru User Manual

Page 36

Advertising
background image

36

8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C

8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

Probes

MKUv 1610

MKUv 1612

H0 = Gradient

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Gradient H0- :

3.31 K

(+2.0 K / –1.31 K from

set temperature)

Max. fluctuation:

2.70 K

Gradient H0- :

4.27 K

(+2.27 K / –2.00 K from

set temperature)

Max. fluctuation:

4.00 K

Liebherr laboratory refrigerators or freezers do not comply with the German Institute for Standardisation norm DIN 58371 regarding the
requirements for the storage of conserved blood and DIN 58375 regarding the requirements for the storage of blood plasma. The MKv and
MKUv models comply with the German Institute for Standardisation norm DIN 58345 which refers to the specification of pharmacy refrigerators.
All other Liebherr refrigerators and freezers in this catalogue do not comply with this standard.

Set temperature

Average probe temperature

Set temperature

Average probe temperature

Pharmacy refrigerators compliant to DIN 58345