Liebherr WKt 4552 GrandCru User Manual

Page 35

Advertising
background image

35

The evaluation of air temperatures of the MK(U)v models according to NF X 15-140 using a PT 100 temperature probe shows the following results:

Results according to NF X 15-140

8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C

8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

Probes

MKv 3910

MKv 3912

H0 = Gradient

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Gradient H0- :

4.20 K

(+2.4 K / –1.8 K from

set temperature)

Max. fluctuation:

3.10 K

Gradient H0- :

4.46 K

(+3.1 K / –1.36 K from

set temperature)

Max. fluctuation:

3.60 K

Set temperature

Average probe temperature

Set temperature

Average probe temperature

Pharmacy refrigerators compliant to DIN 58345