Temperature consistency and distribution
according to NF X 15-140

The Liebherr laboratory cabinets are all tested in-house to the NF X 15-140 test procedure. Liebherr’s modern R & D facility has climate controlled test rooms
which conform to EN 23953.

Description of the NF X 15-140 test procedure

1

2

3

4

6

7

8

9

5

centred

1/10 B

1/10 T

1/10 H

LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller

The general test conditions are as follows:

Ambient temperature + 25 °C

60 % relative humidity

Refrigerator set temperature + 5 °C

Freezer set temperature – 20 °C

Low temperature freezer set temperature – 45 °C

Measurements are made using 9 x PT 100 air temperature probes positioned in accordance with NF X 15-140.

“Unless specifically stipulated, the temperature probes are located at equal distances from the walls of 1/10 of each of the dimensions of the internal volume
(w/h/d). The location of the probes used for characterisation forms a supposedly representative sample of the working space.”

The characterisation is performed using nine probes; one located at each corner, one in the centre of the working space.

Location of the probes used for characterisation