28

9,0 °C
8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C
1,5 °C
1,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

LKv 3912

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

LKv 3910

9,0 °C
8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C
1,5 °C
1,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Gradient H0- :

4.2 K

(+1.8 K / – 2.4 K from

set temperature)

Max. fluctuation:

3.7 K

Gradient H0- :

3.0 K

(+1.25 K / –1.75 K from

set temperature)

Max. fluctuation:

2.5 K

LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller

Set temperature

Average probe temperature

Set temperature

Average probe temperature