Liebherr WKt 4552 GrandCru User Manual

Page 16

Advertising
background image

16

LGPv 1420

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

–17.0 °C

–17.5 °C

–18.0 °C

–18.5 °C

–19.0 °C

–19.5 °C

–20.0 °C

–20.5 °C

–21.0 °C

–21.5 °C

–22.0 °C

–23.0 °C

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Gradient H0- :

3.5 K

(+1.5 K / – 2.0 K from

set temperature)

Max. fluctuation:

3.9 K

LKPv and LGPv laboratory appliances with Profi-Controller

Set temperature

Average probe temperature

LGPv 8420

Gradient H0- :

3.52 K

(+2.0 K / – 1.52 K from

set temperature)

Max. fluctuation:

4.3 K

– 17,0 °C
– 17,5 °C
– 18,0 °C
– 18,5 °C
– 19,0 °C
– 19,5 °C
– 20,0 °C
– 20,5 °C
– 21,0 °C
– 21,5 °C
– 22,0 °C
– 22,5 °C
– 23,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

H0 = Gradient

Probes

Set temperature

Average probe temperature

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty