Liebherr WKt 4552 GrandCru User Manual
Page 16
16
LGPv 1420
t1
t2
t3
t4
t5
t6
t7
t8
t9
Probes
–17.0 °C
–17.5 °C
–18.0 °C
–18.5 °C
–19.0 °C
–19.5 °C
–20.0 °C
–20.5 °C
–21.0 °C
–21.5 °C
–22.0 °C
–23.0 °C
H0 = Gradient
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty
Gradient H0- :
3.5 K
(+1.5 K / – 2.0 K from
set temperature)
Max. fluctuation:
3.9 K
LKPv and LGPv laboratory appliances with Profi-Controller
Set temperature
Average probe temperature
LGPv 8420
Gradient H0- :
3.52 K
(+2.0 K / – 1.52 K from
set temperature)
Max. fluctuation:
4.3 K
– 17,0 °C
– 17,5 °C
– 18,0 °C
– 18,5 °C
– 19,0 °C
– 19,5 °C
– 20,0 °C
– 20,5 °C
– 21,0 °C
– 21,5 °C
– 22,0 °C
– 22,5 °C
– 23,0 °C
t1
t2
t3
t4
t5
t6
t7
t8
t9
H0 = Gradient
Probes
Set temperature
Average probe temperature
max (0
mj
+ U
mj
) = mean value warmest probe + expanded measurement uncertainty
min (0
mj
– U
mj
) = mean value coldest probe – expanded measurement uncertainty