Liebherr WKt 4552 GrandCru User Manual

Page 30

Advertising
background image

30

LK(U)v, LCv, LK(U)exv, LG(U)ex and LGT laboratory appliances with Comfort-Controller

9,0 °C
8,5 °C
8,0 °C
7,5 °C
7,0 °C
6,5 °C
6,0 °C
5,5 °C
5,0 °C
4,5 °C
4,0 °C
3,5 °C
3,0 °C
2,5 °C
2,0 °C
1,5 °C
1,0 °C

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

LCv 4010 – Evaluation of refrigerator compartment

H0 = Gradient

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Gradient H0- :

6.0 K

(+2.6 K / – 3.4 K from

set temperature)

Max. fluctuation:

3.4 K

Gradient H0- :

3.4 K

(+1.8 K / –1.6 K from

set temperature)

Max. fluctuation:

4.0 K

Set temperature

Average probe temperature

LCv 4010 – Evaluation of freezer compartment

t1

t2

t3

t4

t5

t6

t7

t8

t9

Probes

– 16,5 °C
– 17,0 °C
– 17,5 °C
– 18,0 °C
– 18,5 °C
– 19,0 °C
– 19,5 °C
– 20,0 °C
– 20,5 °C
– 21,0 °C
– 21,5 °C
– 22,0 °C
– 22,5 °C
– 23,0 °C
– 23,5 °C
– 24,0 °C

H0 = Gradient

min (0

mj

– U

mj

) = mean value coldest probe – expanded measurement uncertainty

Set temperature

Average probe temperature

max (0

mj

+ U

mj

) = mean value warmest probe + expanded measurement uncertainty